DPS
Drejtoria e Përgjithshme e Standardizimit
Tel/Cel: +355 4 222 62 55
E-mail: info@dps.gov.al
Adresa: Rr.: "Reshit Collaku", (pranë ILDKPKI, kati VI), Kutia Postare 98, Tiranë - Shqipëri
Main menu

DS CR 13825:2000

Lidhësit Luer - Raporti për CEN nga grupi i punës të forumit të CEN “Lidhjet e rakorderive

Luer connectors - A report to CEN chef from the CEN forum task group "Luer fittings"
24 tet 2011

General information

60.60     24 tet 2011

DPS

DPS/KT 47

CEN Report

11.040.20  

anglisht  

Buying

Publikuar

Language in which you want to receive the document.

Scope

Luer connectors are used in medical devices with a wide range of functions, raising the possibility of accidental cross connection between devices. We assessed the hazards arising from accidental misconnection, and reviewed adverse incidents known to have involved Luer connectors.

We considered three possible strategies for reducing the risk of misconnection being made:

1) The use of different male/female sequences for different applications.
2) The use of colour coding.
3) The restriction of the applications for which the use of Luer connectors is
permitted.

Only the last of these offered real benefits.

Related directives

Directives related to this standards.

93/42/EEC

Pajisje Mjekësore

93/42/EEC

Pajisje Mjekësore

Life cycle

NOW

PUBLISHED
DS CR 13825:2000
60.60 Standard published
24 tet 2011

Related project

Adopted from CR 13825:2000

Adopted from CR 13825:2000

Preview

Only informative sections of projects are publicly available. To view the full content, you will need to members of the committee. If you are a member, please log in to your account by clicking on the "Log in" button.

Login