DPS
Drejtoria e Përgjithshme e Standardizimit
Tel/Cel: +355 4 222 62 55
E-mail: info@dps.gov.al
Adresa: Rr.: "Reshit Collaku", (pranë ILDKPKI, kati VI), Kutia Postare 98, Tiranë - Shqipëri
Main menu

SSH EN ISO 10993-6:2016

Vlerësimi biologjik i pajisjeve mjekësore - Pjesa 6: Testet për efektet lokale pas implantimit (ISO 10993-6:2016)

Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2016)
23 tet 2017

General information

60.60     17 korr 2017

DPS

DPS/KT 206

European Norm

11.100.20  

anglisht  

Buying

Publikuar

Language in which you want to receive the document.

Scope

ISO 10993-6:2016 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.
ISO 10993-6:2016 applies to materials that are
- solid and non-absorbable,
- non-solid, such as porous materials, liquids, gels, pastes, and particulates, and
- degradable and/or absorbable, which may be solid or non-solid.
The test sample is implanted into a site and animal species appropriate for the evaluation of the biological safety of the material. These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical or functional loading. This part of ISO 10993 can also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining might have been breached, in order to evaluate local tissue responses.
The local effects are evaluated by a comparison of the tissue response caused by a test sample to that caused by control materials used in medical devices whose clinical acceptability and biocompatibility characteristics have been established. The objective of the test methods is to characterize the history and evolution of the tissue response after implantation of a medical device/biomaterial including final integration or absorption/degradation of the material. In particular for degradable/absorbable materials, the degradation characteristics of the material and the resulting tissue response should be determined.
ISO 10993-6:2016 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects might provide insight into some of these properties. Systemic toxicity studies conducted by implantation might satisfy the requirements of this part of ISO 10993. When conducting combined studies for evaluating local effects and systemic effects, the requirements of both standards is to be fulfilled.

Related directives

Directives related to this standards.

90/385/EEC

Pajisjet mjekësore aktive transplatuese

93/42/EEC

Pajisje Mjekësore

2007/47/EC

Direktiva 2007/47/EC e Parlamentit Europian dhe e Këshillit të 5 shtator 2007 amendon Direktivën e Këshillit 90/385/EEC mbi përafrimin e ligjeve të Shteteve Anëtare në lidhje me pajisje medicinale aktive transplatuese, Direktiva e Këshillit 93/42/EEC në lidhje me pajisjet mjekësore dhe Direktivën 98/8/EC në lidhje me vendosjen e produkteve biocide në treg

Life cycle

PREVIOUSLY

WITHDRAWN
SSH EN ISO 10993-6:2009

NOW

PUBLISHED
SSH EN ISO 10993-6:2016
60.60 Standard published
17 korr 2017

Related project

Adopted from EN ISO 10993-6:2016

Adopted from ISO 10993-6:2016