DPS
Drejtoria e Përgjithshme e Standardizimit
Phone: +355 4 222 62 55
E-mail: info@dps.gov.al
Address: Address: "Reshit Collaku" Str., (nearby ILDKPKI, VI floor), Po.Box 98, Tiranë - Albania
Main menu

SSH EN ISO 10993-6:2016

Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2016)

Oct 23, 2017

General information

60.60     Jul 17, 2017

DPS

DPS/KT 206

European Norm

11.100.20  

English  

Buying

Published

Language in which you want to receive the document.

Scope

ISO 10993-6:2016 specifies test methods for the assessment of the local effects after implantation of biomaterials intended for use in medical devices.
ISO 10993-6:2016 applies to materials that are
- solid and non-absorbable,
- non-solid, such as porous materials, liquids, gels, pastes, and particulates, and
- degradable and/or absorbable, which may be solid or non-solid.
The test sample is implanted into a site and animal species appropriate for the evaluation of the biological safety of the material. These implantation tests are not intended to evaluate or determine the performance of the test sample in terms of mechanical or functional loading. This part of ISO 10993 can also be applied to medical devices that are intended to be used topically in clinical indications where the surface or lining might have been breached, in order to evaluate local tissue responses.
The local effects are evaluated by a comparison of the tissue response caused by a test sample to that caused by control materials used in medical devices whose clinical acceptability and biocompatibility characteristics have been established. The objective of the test methods is to characterize the history and evolution of the tissue response after implantation of a medical device/biomaterial including final integration or absorption/degradation of the material. In particular for degradable/absorbable materials, the degradation characteristics of the material and the resulting tissue response should be determined.
ISO 10993-6:2016 does not deal with systemic toxicity, carcinogenicity, teratogenicity or mutagenicity. However, the long-term implantation studies intended for evaluation of local biological effects might provide insight into some of these properties. Systemic toxicity studies conducted by implantation might satisfy the requirements of this part of ISO 10993. When conducting combined studies for evaluating local effects and systemic effects, the requirements of both standards is to be fulfilled.

Related directives

Directives related to this standards.

90/385/EEC

Active implantable medical devices

93/42/EEC

Medical devices

2007/47/EC

Directive 2007/47/EC of the European Parliament and of the council of 5 September 2007 amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market

Life cycle

PREVIOUSLY

WITHDRAWN
SSH EN ISO 10993-6:2009

NOW

PUBLISHED
SSH EN ISO 10993-6:2016
60.60 Standard published
Jul 17, 2017

Related project

Adopted from EN ISO 10993-6:2016

Adopted from ISO 10993-6:2016