ISO 10993‑13:2010 provides general requirements for the design of tests in a simulated environment for identifying and quantifying degradation products from finished polymeric medical devices ready for clinical use.
ISO 10993‑13:2010 describes two test methods to generate degradation products, an accelerated degradation test as a screening method and a real-time degradation test in a simulated environment. For materials that are intended to polymerize in situ, the set or cured polymer is used for testing. The data generated are used in the biological evaluation of the polymer. ISO 10993‑13:2010 considers only non-resorbable polymers. Similar, but appropriately modified procedures may be applicable for resorbable polymers.
ISO 10993‑13:2010 considers only those degradation products generated by a chemical alteration of the finished polymeric device. It is not applicable to degradation of the device induced during its intended use by mechanical stress, wear or electromagnetic radiation or biological factors such as enzymes, other proteins and cellular activity.
Directives related to this standards.
Direktiva 2007/47/EC e Parlamentit Europian dhe e Këshillit të 5 shtator 2007 amendon Direktivën e Këshillit 90/385/EEC mbi përafrimin e ligjeve të Shteteve Anëtare në lidhje me pajisje medicinale aktive transplatuese, Direktiva e Këshillit 93/42/EEC në lidhje me pajisjet mjekësore dhe Direktivën 98/8/EC në lidhje me vendosjen e produkteve biocide në treg