SSH EN 61558-2-15:2001

Siguria e transformatorëve të forcës. blloqeve të furnizimit të ushqimit dhe të ngjashëm me ta - Pjesa 2 - 15: Kërkesa të veçanta për transformatorët izolues për pajisjet e lokaleve mjekësore

Safety of power transformers, power supply units and similar - Part 2-15: Particular requirements for isolating transformers for the supply of medical locations
13 maj 2008
95.99 Withdrawal of Standard   19 jan 2017

General information

95.99     19 jan 2017

DPS

DPS/KT 4

European Norm

29.180  

anglisht  

Buying

Shfuqizuar

Language in which you want to receive the document.

Scope

This part 2-15 of IEC 61558 applies to stationary, single-phase or polyphase, air-cooled (natural or forced) isolating transformers for the supply of group II medical locations, designed to be permanently connected to the fixed wiring of IT supply system. This standard also applies to transformers incorporating electronic circuits. This standard does not apply to external circuits and their components intended to be connected to the input and output terminals or socket-outlets of the transformer. It has the status of a group safety publication in accordance with IEC Guide 104. This part 2 is intended to be used in conjunction with IEC 61558-1.

Related directives

Directives related to this standards.

2006/95/EC

Direktiva 2006/95/EC e Parlamentit Europian dhe e Këshillit e 12 dhjetorit 2006 mbi harmonizimin e ligjeve të Shteteve anëtare lidhur me pajisjet elektrike të projektuara për përdorim brenda disa kufijve të tensionit të caktuar (versioni i kodifikuar)

Të harmonizuara

Life cycle

NOW

WITHDRAWN
SSH EN 61558-2-15:2001
95.99 Withdrawal of Standard
19 jan 2017

REVISED BY

PUBLISHED
SSH EN 61558-2-15:2012

Related project

Adopted from EN 61558-2-15:2001

Adopted from IEC 61558-2-15 Ed. 1.0 b