ISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator.
ISO 11137-1:2006 does not:
specify requirements for development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease;
detail specified requirements for designating a medical device as sterile;
specify a quality management system for the control of all stages of production of medical devices;
specify requirements for occupational safety associated with the design and operation of irradiation facilities;
specify requirements for the sterilization of used or reprocessed devices.
Directives related to this standards.
WITHDRAWN
SSH EN ISO 11137-1:2007
95.99
Withdrawal of Standard
30 mar 2016
PUBLISHED
SSH EN ISO 11137-1:2015