ISO 19996:2024

Charge conditioning of aerosol particles for particle characterization and the generation of calibration and test aerosols

Oct 18, 2024

General information

60.60     Oct 18, 2024

ISO

ISO/TC 24/SC 4

International Standard

19.120  

English  

Buying

Published

Language in which you want to receive the document.

Scope

This document specifies requirements and provides guidance for the use of charge conditioners for aerosol particles, especially for particle characterization and for the generation of calibration and test aerosols.
This document provides a methodology to specify the performance of charge conditioners and for adequate quality control, with respect to their application in:
— particle size and concentration measurement with differential mobility analysing systems (DMAS);
—particle size classification with differential electrical mobility classifiers (DEMC).
For these applications, this document covers particle charge conditioning for particle sizes ranging from approximately 1 nm to 1 µm and for particle number concentrations at the inlet of the charge conditioner up to approximately 107 cm-3.
This document does not address specific charge conditioner designs or other applications besides those specified in Clause 1.
Radiation safety for charge conditioners with radioactive sources or x-ray tubes is not covered by this document.