DPS
Drejtoria e Përgjithshme e Standardizimit
Tel/Cel: +355 4 222 62 55
E-mail: info@dps.gov.al
Adresa: Rr.: "Reshit Collaku", (pranë ILDKPKI, kati VI), Kutia Postare 98, Tiranë - Shqipëri
Main menu

ISO/TR 18130:2016

Dentistry — Screw loosening test using cyclic torsional loading for implant body/implant abutment connection of endosseous dental implants
19 maj 2016

General information

60.60     19 maj 2016

ISO

ISO/TC 106/SC 8

Technical Report

11.060.15  

frëngjisht   anglisht  

Buying

Publikuar

Language in which you want to receive the document.

Scope

ISO/TR 18130:2016 provides guidelines for a method to determine the extent of screw loosening of the metallic implant body/implant abutment joint of endosseous dental implants, such as two-part implants or multi-part implants under cyclic torsional loading. This test is most appropriate for evaluating new types of joints fixed using screw(s) and metallic connecting parts. This Technical Report provides a protocol for torsional cyclic torque on an implant body/implant abutment joint, but its intended use is for prefabricated implant bodies, implant abutments and, if appropriate, implant connecting parts that are made of metallic materials.
It is not applicable to ensure the in vivo performance of endosseous dental implants and is not derived from observations of clinical failures.
NOTE This Technical Report is not intended for use with temporary abutments.

Life cycle

NOW

PUBLISHED
ISO/TR 18130:2016
60.60 Standard published
19 maj 2016

National adoptions

Stomatologji - Testi i lirimit të vidhave duke përdorur ngarkesën rrotulluese ciklike për lidhjen e trupit të implantit/mbështjellësit të implanteve dentare endoseoze

60.60 Standard published

DPS/KT 55 më tepër

Preview

Only informative sections of projects are publicly available. To view the full content, you will need to members of the committee. If you are a member, please log in to your account by clicking on the "Log in" button.

Login