IEC 60601-2-29:2008 ED3

Medical electrical equipment - Part 2-29: Particular requirements for the basic safety and essential performance of radiotherapy simulators
11 qer 2008

General information

60.60     11 qer 2008

IEC

TC 62/SC 62C

International Standard

11.040.60  

anglisht   frëngjisht   spanjisht  

Buying

Publikuar

Language in which you want to receive the document.

Scope

IEC 60601-2-29:2008 establishes requirements to be complied with by manufacturers in the design and construction of radiotherapy simulators; it does not attempt to define their optimum performance requirements. Its purpose is to identify those features of design that are regarded, at the present time, as essential for the safe operation of such medical electrical equipment. It places limits on the degradation of medical electrical equipment performance beyond which it can be presumed that a fault condition exists, for example a component failure, and where an interlock then operates to prevent continued operation of the medical electrical equipment. This third edition cancels and replaces the second edition published in 1999. This edition constitutes a technical revision, which brings this standard in line with the third edition of IEC 60601-1 and its collateral standards.

Life cycle

PREVIOUSLY

WITHDRAWN
IEC 60601-2-29:1999 ED2

NOW

PUBLISHED
IEC 60601-2-29:2008 ED3
60.60 Standard published
11 qer 2008