Revised
This International Standard
- specifies a method of determining the total emitted acoustic power of ultrasonic transducers based on the use of a radiation force balance;
- establishes general principles for the use of radiation force balances in which an obstacle (target) intercepts the sound field to be measured;
-establishes limitations of the radiation force method related to cavitation and temperature rise;
- establishes quantitative limitations of the radiation force method in relation to diverging and focused beams;
- provides information on assessment of overall measurement uncertainties.
This International Standard is applicable to:
- the measurement of ultrasonic power up to 1 W based on the use of a radiation force balance in the frequency range from 0,5 MHz to 25 MHz;
- the measurement of ultrasonic power up to 20 W based on the use of a radiation force balance in the frequency range 0,75 MHz to 5 MHz;
- the measurement of total ultrasonic power of transducers, preferably with well-collimated beams;
- the use of radiation force balances of the gravimetric type or force feedback type.
WITHDRAWN
IEC 61161:2006 ED2
99.60
Withdrawal effective
30 jan 2013
PUBLISHED
IEC 61161:2013 ED3