IEC 61010-2-033:2019 ED2

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-033: Particular requirements for hand-held multimeters and other meters for domestic and professional use, capable of measuring mains voltage
21 qer 2019

General information

99.60     20 sht 2023

IEC

TC 66

International Standard

19.080     71.040.10  

anglisht  

Buying

Revised

Language in which you want to receive the document.

Scope

IEC 61010-2-033:2019 specifies safety requirements for hand-held multimeters for domestic and professional use, capable of measuring mains. Hand-held multimeters are multi-range multifunction measuring instruments intended to measure voltage and other electrical quantities such as resistance or current. Their primary purpose is to measure voltage on a live mains. They are suitable to be supported by one hand during normal use. This second edition cancels and replaces the first edition published in 2012. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- The scope has been reduced to hand-held multimeters. Voltmeters and clamp multimeters have been removed. They are addressed respectively by IEC 61010-2-030 and IEC 61010-2-032. The relevant definitions have been removed.
- Subclause 4.4.2.101 has been relocated into Clause 102.
- Clearances and creepage distances for wet locations and for measuring circuit terminals exceeding 1 000 V a.c. or 1 414 V d.c. have been specified.
- Subclause 14.101 related to "Circuits or components used as transient overvoltage limiting devices in measuring circuits used to measure mains" has been removed.
- References to IEC 61010-031 for probe assemblies and IEC 61010-2-032 for current sensors have been added.
- Requirements for protection against mains overvoltage measuring circuits have been added.
- Clause 102 has been rewritten.
- Requirements for measuring circuits from 1 000 V to 3 000 V have been added.
- An informative Annex CC about dimensions of 4-mm banana terminals has been added.
- A flowchart for insulation according to the type of circuit has been added in a new Annex DD.

Life cycle

PREVIOUSLY

WITHDRAWN
IEC 61010-2-033:2012 ED1

NOW

WITHDRAWN
IEC 61010-2-033:2019 ED2
99.60 Withdrawal effective
20 sht 2023

REVISED BY

PUBLISHED
IEC 61010-2-033:2023 ED3

National adoptions

Kërkesa të sigurisë për pajisjet elektrike për matje, kontroll dhe përdorim në laborator - Pjesa 2-033: Kërkesat të veçanta për multimetrat e dorës dhe pajisjeve matëse të tjera, për përdorim shtëpiak dhe profesional, të aftë për të matur tensionin e rrjetit

95.99 Withdrawal of Standard