IEC 60645-7:2009 applies to instruments designed for the measurement of auditory evoked potentials from the inner ear, the auditory nerve and the brainstem, evoked by acoustic and/or vibratory stimuli of short duration. IEC 60645-7:2009 defines the characteristics to be specified by the manufacturer, specifies performance requirements for two types of instrument, screening and diagnostic, and specifies the functions to be provided on these types. The purpose of IEC 60645-7:2009 is to ensure that measurements made under comparable test conditions with different instruments complying with this standard will be consistent. IEC 60645-7:2009 is not intended to restrict development or incorporation of new features, nor to discourage innovative approaches. The application of electric stimuli for special purposes is beyond the scope of this standard.
Directives related to this standards.
Direktiva 2006/95/EC e Parlamentit Europian dhe e Këshillit e 12 dhjetorit 2006 mbi harmonizimin e ligjeve të Shteteve anëtare lidhur me pajisjet elektrike të projektuara për përdorim brenda disa kufijve të tensionit të caktuar (versioni i kodifikuar)
Të harmonizuaraDirektiva e BE 2014/35 e Parlamentit Europian dhe e Këshillit e datës 26 Shkurt 2014 mbi harmonizimin e ligjeve të Shteteve Anëtare lidhur me vënien në dispozicion të tregut të pajisjeve elektrike të destinuara për përdorim brenda disa kufijve të caktuar të tensionit
Të harmonizuara
PUBLISHED
SSH EN 60645-7:2010
60.60
Standard published
1 jan 2010
IN_DEVELOPMENT
prSSH EN IEC 60645-7:2025