DPS
Drejtoria e Përgjithshme e Standardizimit
Phone: +355 4 222 62 55
E-mail: info@dps.gov.al
Address: Address: "Reshit Collaku" Str., (nearby ILDKPKI, VI floor), Po.Box 98, Tiranë - Albania
Main menu

SSH EN IEC 60318-8:2022

Electroacoustics - Simulators of human head and ear - Part 8: Acoustic coupler for high-frequency measurements of hearing aids and earphones coupled to the ear by means of ear inserts

Jul 12, 2023

General information

60.60     Jul 12, 2023

95.99   

DPS

DPS/KT 8

European Norm

17.140.50  

English  

Buying

Published

Language in which you want to receive the document.

Scope

IEC 60318-8:2022 describes an acoustic coupler for loading a hearing aid or insert earphone with a specified acoustic impedance when testing its acoustic performance, in the frequency range up to 16 kHz. It is suitable for air-conduction hearing aids and earphones, coupled to the ear by means of ear inserts, earmoulds or similar devices.

The acoustic coupler does not simulate the human ear. However, it has an effective volume of only 0,4 cm<sup>3</sup>, which is small enough not to produce significant resonances in the coupler in the frequency range below 16 kHz. Therefore, it will load the earphone with a known acoustic impedance, which allows repeatable measurements with low uncertainty to be obtained on earphones used in extended high-frequency audiometry.

Life cycle

NOW

PUBLISHED
SSH EN IEC 60318-8:2022
60.60 Standard published
Jul 12, 2023

Related project

Adopted from EN IEC 60318-8:2022

Adopted from IEC 60318-8:2022