ISO 16063-11:1999

Methods for the calibration of vibration and shock transducers — Part 11: Primary vibration calibration by laser interferometry

Dec 16, 1999

General information

90.20     Jan 15, 2026

ISO

ISO/TC 108

International Standard

17.160  

English  

Buying

Published

Language in which you want to receive the document.

Scope

This part of ISO 16063 specifies the instrumentation and procedure to be used for primary vibration calibration of
rectilinear accelerometers (with or without amplifier) to obtain magnitude and phase lag of the complex sensitivity
by steady-state sinusoidal vibration and laser interferometry.
It is applicable to a frequency range from 1 Hz to 10 kHz and a dynamic range (amplitude) from 0,1 m/s2 to
1 000 m/s2 (frequency-dependent).
These ranges are covered with the uncertainty of measurement specified in clause 2. Calibration frequencies lower
than 1 Hz (e.g. 0,4 Hz, which is a reference frequency used in other International Standards) and acceleration
amplitudes smaller than 0,1 m/s2 (e.g. 0,004 m/s2 at 1 Hz) can be achieved using Method 3 specified in this part of
ISO 16063, in conjunction with an appropriate low-frequency vibration generator.
Method 1 (fringe-counting method) is applicable to sensitivity magnitude calibration in the frequency range 1 Hz to
800 Hz and, under special conditions, at higher frequencies (cf. clause 7). Method 2 (minimum-point method) can
be used for sensitivity magnitude calibration in the frequency range 800 Hz to 10 kHz (cf. clause 8). Method 3
(sine-approximation method) can be used for magnitude of sensitivity and phase calibration in the frequency range
1 Hz to 10 kHz (cf. clause 9).
Methods 1 and 3 provide for calibrations at fixed acceleration amplitudes at various frequencies. Method 2 requires
calibrations at fixed displacement amplitudes (acceleration amplitude varies with frequency).

Life cycle

PREVIOUSLY

WITHDRAWN
ISO 5347-1:1993

NOW

PUBLISHED
ISO 16063-11:1999
90.20 Standard under periodical review
Jan 15, 2026

REVISED BY

IN_DEVELOPMENT
ISO/PWI 16063-11