DPS
Drejtoria e Përgjithshme e Standardizimit
Phone: +355 4 222 62 55
E-mail: info@dps.gov.al
Address: Address: "Reshit Collaku" Str., (nearby ILDKPKI, VI floor), Po.Box 98, Tiranë - Albania
Main menu

IEC 62127-2 ED2

Ultrasonics – Hydrophones – Part 2: Calibration for ultrasonic fields

General information

60.00     Nov 1, 2024

PPUB    Dec 13, 2024

IEC

TC 87

International Standard

17.140.50  

Scope

IEC 62127-2:2024 specifies:
• absolute hydrophone calibration methods;
• relative (comparative) hydrophone calibration methods.
Recommendations and references to accepted literature are made for the various relative and absolute calibration methods in the frequency range covered by this document.
This document is applicable to
• hydrophones used for measurements made in water and in the ultrasonic frequency range 50 kHz to 100 MHz;
• hydrophones employing piezoelectric sensor elements, designed to measure the pulsed wave and continuous wave ultrasonic fields generated by ultrasonic equipment;
• hydrophones with or without a hydrophone pre-amplifier.
IEC 62127-2:2024 cancels and replaces the first edition published in 2007, Amendment 1:2013 and Amendment 2:2017. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) the upper frequency limit of 40 MHz has been removed;
b) hydrophone sensitivity definitions have been changed to recognize sensitivities as complex-valued quantities;
c) directional response measurement and effective size determination procedures have been updated in 12.5.1 to align with recent changes in IEC 62127-3;
d) Annex F has been amended to comprise a calibration technique for high-frequency complex-valued calibration;
e) the reciprocity method description in Annex K was extended to also comprise focusing transducers;

Life cycle

NOW

IN_DEVELOPMENT
IEC 62127-2 ED2
60.00 Standard under publication
Nov 1, 2024