IEC 61010-2-040:2020 ED3

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-040: Particular requirements for sterilizers and washer-disinfectors used to treat medical materials

May 14, 2020

General information

60.60     May 14, 2020

IEC

TC 66

International Standard

19.080     71.040.10  

English  

Buying

Published

Language in which you want to receive the document.

Scope

IEC 61010-2-040:2020 specifies safety requirements for electrical equipment intended for sterilization, washing, and disinfection of medical materials in the medical, veterinary, pharmaceutical and laboratory fields, when used under the environmental conditions of 1.4. Examples of such equipment include the following:
- sterilizers and disinfectors using steam and/or hot water as the sterilant;
- sterilizers and disinfectors using toxic gas, toxic aerosol or toxic vapour as the sterilant;
- sterilizers and disinfectors using hot air or hot inert gas as the sterilant; and
- washer-disinfectors.
This third edition cancels and replaces the second edition published in 2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- it is established on the basis of the third edition (2010) of IEC 61010‑1 and its Amendment 1 (2016);
- added tolerance for stability of a.c. voltage test equipment to 6.8.3.1;
- the status of a Group Safety Publication has been removed (this does not change the technical requirements in the document).

Life cycle

PREVIOUSLY

WITHDRAWN
IEC 61010-2-040:2015 ED2

NOW

PUBLISHED
IEC 61010-2-040:2020 ED3
60.60 Standard published
May 14, 2020

REVISED BY

IN_DEVELOPMENT
IEC 61010-2-040 ED4

National adoptions

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-040: Particular requirements for sterilizers and washer-disinfectors used to treat medical materials

60.60 Standard published