SSH EN 62129-1:2016

Calibration of wavelength/optical frequency measurement instruments - Part 1: Optical spectrum analyzers

Mar 21, 2017

General information

60.60     Jan 26, 2017

DPS

DPS/KT 3

European Norm

33.140     33.180.01  

English  

Buying

Published

Language in which you want to receive the document.

Scope

IEC 62129-1:2016 specifies procedures for calibrating an optical spectrum analyzer that is developed for use in fibre-optic communications and designed to measure the power distribution of an optical spectrum. It does not apply to an optical wavelength meter that measures only centre wavelengths, a Fabry-Perot interferometer or a monochromator that has no display unit. This first edition of IEC 62129-1 cancels and replaces the first edition of IEC 62129, published in 2006. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:

- update of term and definitions;

- update of calibration conditions;

- calculation change of uncertainties related to wavelength temperature dependence, power linearity, power level temperature dependence;

- move of Annex E to the bibliography.

Keywords: calibrating an optical spectrum analyser

Life cycle

PREVIOUSLY

WITHDRAWN
SSH EN 62129:2006

NOW

PUBLISHED
SSH EN 62129-1:2016
60.60 Standard published
Jan 26, 2017

Related project

Adopted from EN 62129-1:2016

Adopted from IEC 62129-1 Ed. 1.0 b